Diagnostic Software Utilities
5-36
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
Procedure Details
6.
On the acquisition PC’s desktop, right-
click the My Computer icon and select
Manage from the pop-up menu.
The Computer Management window appears.
7.
In the left-hand pane, select Device
Manager.
8.
In the right-hand pane, click the + next to
the entry Ports (Com & LPT).
9.
Check the COM port number stated next
to the item ATEN to USB Serial Bridge.
On a tower acquisition PC, you must use
COM1.
10.
Ensure that the correct COM port is
selected.
11.
Click OK to continue.
The COM Properties dialog appears.
12.
Set the following parameters for the serial
interface:
Baud rate: 57 600
Data bits: 8
Parity: None
Stop bits: 1
Flow control: Hardware.
13.
Click OK to continue.
III
III
I, II
III
I, II