Gantry
3-16
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
3.2.4
Motor Amplifiers
Figure 3.11 Motion controllers and motor amplifiers inside table console
Figure 3.11 shows the motor amplifiers.
Note
. The X amplifier has a built-in potentiometer, which is adjusted correctly when the
CardioMD is delivered from the manufacturer. Changing the setting of this potentiometer does not
affect CardioMD application. Therefore, do
not
change the setting of the X amplifier’s
potentiometer.
3.2.5
Connections, Series III Systems
3.2.5.1
Rear Panel
CardioMD Series III systems have a connector panel on the rear side of the table console. See
Figure 3.12 on page 3-17.
ECG Gate Power Supply
An ECG gate can be supplied from this 115 V AC output. The output is protected by a circuit
breaker. A green LED in the panel shows the status of the 115 V output, which also supplies the
acquisition PC.
ECG Trigger
The ECG Trigger connector is used to connect the R-wave trigger signal from the ECG gate. The
gate is connected to the detector EDC boards, where the trigger pulses are detected and embedded
in the FireWire event stream transmitted to the acquisition PC as time references used for the
image framing process.
X amplifier
Rotate amplifier
Y amplifier
Rotate motion
controller
X/Y motion
controller
III
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......