CardioMD AC Option
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
12-97
Procedure Details
9.
Using the 4 pcs. M4 socket-head cap
screws provided (2SCR0223), mount the
AC Option power supply (9POW1315)
inside the table console frame on the four
stand-offs just mounted.
10.
Locate the orange-colored, wire-mounted
4-way connector plug inside the table
console.
11.
Plug it into the orange-colored, wire-
mounted 4-way connector socket attached
to the AC Option power supply.