Installation
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
8-39
8.3.3.21
Electrical Connections between Detector, X-drive, Y-drive and Table Console
Procedure Details
1.
For all three cable conduits, unscrew the
nuts and place the cables in the three
slotted holes at the bottom of the table
console.
The thick conduit from the detector rotate gear
goes into the right-hand hole.
The conduit from the gantry base goes into the
middle hole.
The conduit from the bottom of the Y-pillar
goes into the left hole.
2.
Tighten the nuts by hand.
3.
From the front side of the table console,
remove the fixing screws of the two
rectangular ferrite filters in the front
compartment of the table console.