Installation
8-38
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
8.3.3.18
Leveling the Table Base
Procedure Details
1.
Remove the mounting screws in the rear
end of the table base.
2.
Using a level and the adjusting tool, level
the table base.
3.
Remount the mounting screws.
8.3.3.19
Removing the 2 Shipping Locks
Procedure Details
1.
Tighten the mounting screws connecting
the table base and gantry frame.
2.
Using an 6 mm Allen wrench, remove the
2 shipping locks located in the gantry
sidebars.
8.3.3.20
Mounting the Patient Table Support
Procedure Details
1.
Mount the table end support onto the
gantry base using two M8 bolts.