Diagnostics
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
5-45
Procedure Details
9.
On the Persistence page, click the Tools
button.
10.
When the Calibration & Configuration
menu appears, click Service.
11.
If prompted, enter the service password
(currently DDD).
12.
In the Service Calibrations menu, click
Node configuration.
13.
In the Nodes Configuration dialog,
double-click the
+
to add a new node.
The Add a new node dialog appears.
14.
Select the FOLDER option and click
Select.
15.
Check off import/export capabilities from
the node as shown: Yes to Export and No
to Import.
Note
. If you want to both import and export
from the node, check off Yes in both
selections.
16.
To select the folder to where files will be
exported, click the … button.
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......