Diagnostic Software Utilities
5-40
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
5.6.3
FireViewer
This program is part of the FireWire driver installation on the acquisition PC. The program
enables you to perform a quick check of the integrity of the FireWire communication.
Procedure Details
To access the program:
1.
Click the Windows Start button, and then
select
Program
,
ubCore
,
FireViewer
.
The FireViewer program is located in the
directory
c:\Program
Files\Unibrain\ubCore
.
When the FireViewer program has been
started, it immediately shows the status of the
communication.
In CardioMD Series III systems, the FireWire
comprises four nodes:
•
The controlling node, i.e. the acquisition
PC, shown as a white box
•
The FireWire hub shown as a hatched box
(normal operation)
•
The two regular nodes, i.e. the two
detector EDC modules, shown as green
boxes.
The FireViewer window shown in the example
indicates normal communication.