Repair Procedures
12-76
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
Procedure Details
2.
Use a 2.5 mm Allen key to remove the
two screws holding the attenuator.
3.
Pull out the attenuator in the direction of
the arrows.
Caution
. The attenuator consists of four separate copper strips. Be careful to
retract all four strips, aligning the holes of each strip exactly with the holes in
the line source housing.
Procedure Details
4.
Remount the two screws to lock the
attenuator in the open position.
The photos on the right show the attenuator in
the open position.
5.
Reinstall the line source housing inside the
line source compartment. See instructions
in the section
Installing a New Line
Source
, page 12-71.
6.
After reinstalling the line source housing,
grease the acme screw inside the line
source compartment. See section 12.8.1,
page 12-116 for instructions.
7.
Go through the procedure described in
section 12.6.5, page 12-109 to ensure that
the CardioMD system is ready for use.
CAUTION
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......