Installation
8-40
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
Procedure Details
4.
Pull the CANbus cable (coming out of the
thick conduit with a DB9 connector)
through one of the openings at the
upper
edge of the separating wall in the console.
5.
Plug the DB9 into the connector labeled
BZ3 on the motion controller to the right.
6.
Plug the DB15-HD into the mating
connector on the rotate motor amplifier
(the black unit to the right).
7.
Tighten the retaining screws.
8.
Plug the DB15-HD into the mating
connector on the vertical motor amplifier
(the black unit in the middle).
9.
Tighten the retaining screws.
10.
Attach the rectangular ferrite filters to the
standoffs where they came off, using the
fixing screws.
11.
Go to the back of the table console.
12.
Plug the two connectors labeled C3 and
C4 into two similarly labeled connectors
on the power supply panel.
13.
Attach the eye terminal of the
yellow/green wire to the ground screw on
the power supply frame.
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......