Diagnostics
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
5-55
Procedure Details
11.
In the Telnet Properties dialog, select
Automatic from the drop-down list Startup
type.
12.
Click Apply.
13.
Click Start.
14.
Click OK.
The Telnet server is now set up and running
on the acquisition PC.
5.8.2
Windows XP
If your acquisition PC is running Windows XP, FTP and Telnet are already installed. All you need
to do is to configure the Telnet server.
Procedure Details
1.
In the Windows Start menu, select
Run
.
2.
In the Run dialog, enter
cmd
and click
OK.
A command window appears.
3.
At the
>
prompt, enter the command
tnltadmn config sec –ntlm
+passwd [Enter].
5.
Click the Close button in the upper right
corner of the window to close the
command window.
6.
In the Windows Start menu, select
Settings
,
Control Panel
,
Administrative
Tools
,
Services
.
III