14.
Loosen the 22 mm nut holding the
solenoid.
15.
Replace the solenoid, and mount the
22 mm nut.
16.
Loosen the nut and bolt holding the anchor
to the bracket.
17.
Replace the anchor.
Note
. Do
not
tighten the nut and bolt holding
the anchor. It is important that the anchor is
held quite loosely to the bracket.
Note
. It is important that the anchor is
replaced along with the solenoid.
18.
Mount the side rail onto the line source
compartment.
Caution
. When mounting the side rail, take care not to damage the source
motion board.
CAUTION