Repair Procedures
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
6-11
Procedure Details
9.
Make sure that the new crystal has a
gasket placed on its seat.
10.
Put down the new crystal and connect the
HV cable.
11.
Place a GeoCal calibration mask in the
collimator’s place.
12.
Position the crystal so that the holes in the
GeoCal mask line up exactly with the
corresponding marks on the front of the
crystal.
Note
. The print on the crystal cover may not
be entirely centered. If this is the case, position
the crystal so that the marks on the crystal
front are off-centered equally with the holes in
the GeoCal Mask.
13.
Mount the two crystal support brackets,
making sure that no cables get pinched
under the brackets. Tighten the set screws
in the brackets.
14.
Put down the mu-metal assembly and
center it relative to the crystal.
15.
Assemble the detector, reversing the
procedure described above.