CardioMD AC Option
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
12-91
Procedure Details
6.
Take out one transmission scanner, and
then the other, from inside the box.
You are now ready to start the installation of
the CardioMD AC option as described in the
following sections.
WARNING
Be careful when handling the transmission scanners – they are quite heavy. Do
not
drop a transmission scanner. Dropping a transmission scanner may cause
personal injury or damage the equipment.
12.6.2
Electrical Installation
12.6.2.1
Modifying Detector Cabling
Procedure Details
1.
Follow the procedure provided in
Chapter 3
Getting Started
of the
CardioMD Operator’s Manual to shut
down the system.
2.
When the system is powered down,
disconnect the mains cable from the mains
inlet on the rear of the table console.
3.
Remove the detector top cover.
4.
Disconnect the input power connector
from the detector power supply.