CardioMD AC Option
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
12-81
Procedure Details
4.
Again, press down each collision sensor at
the other end of the line source
compartment, one by one, easing the cover
into place.
5.
Remount the edge cover and fasten it
using the two screws previously removed
from the service cover.
6.
Repeat the procedures outlined above for
the other transmission scanner.
12.5.3.11
Mounting Transmission Scanners
Instructions for mounting the transmission scanners on the detectors are provided in section
12.6.3.4 on page 12-103.
12.5.3.12
Aligning Transmission Scanners
The AC option transmission scanners must be aligned:
•
When the AC option is installed for the first time
•
Whenever a line source compartment has been detached from the AC option bracket
•
Whenever a line source housing has been removed from the line source compartment
•
After installation of new line sources
•
When the CardioMD system issues an error message saying “Line source offset has exceeded
+/– X mm”.
Note
. Aligning transmission scanners is performed with the transmission scanners mounted on the
detectors.
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......