CardioMD AC Option
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
12-51
Procedure Details
27.
Fasten the collision sensor to the end of
the line source compartment, mounting
one screw. See photo on the right.
28.
Assemble the spring and end stop as
shown in the photo.
29.
Attach the spring and end stop to the
collision detection shutter end cover.
30.
Mount the collision detection shutter end
cover onto the line source compartment.
Remember to mount the nut on the inside of
the line source compartment behind the spring
and end stop.
Proceed to remount covers, assemble the
transmission scanner and install it on the
detector:
31.
Install the cable connecting the line source
compartment to the bracket by reversing
Summary of Contents for CARDIOMD