Acquisition PC Software Updates
7-18
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
7.3
Acquisition PC Software Updates
The CardioMD software CD-ROM includes two batch files that are used to update or completely
install the acquisition software.
7.3.1
Upgrading CardioMD Acquisition Software
7.3.1.1
Tools Required
•
CardioMD software installation CD, part number 9CSY0906.
Note
. As a precaution, in case anything should go wrong, ensure that the previous version of the
software CD-ROM is available in order to be able to restore the original acquisition software ver-
sion.
7.3.1.2
Procedure
Before starting the acquisition PC software update, check the release note on the CD-ROM. You
find the release note in the directory
Release notes
. The release note may provide information
that is more recent than this manual.
Procedure Details
Before starting the acquisition software up-
grade, close down the CardioMD application:
1.
If the CardioMD application is not dis-
playing the Persistence page, click the Exit
button to return to the Persistence page.
2.
On the Persistence page, click the Close
button in the upper right corner of the
CardioMD application window.
3.
Close all other programs running on the
acquisition PC.
4.
Insert the CardioMD software CD in the
acquisition PC’s CD-ROM drive.
5.
Start the Windows Explorer and select the
CD-ROM drive.
6.
Run the file
InstallCardioMDup-
date.bat
by double-clicking on the
name of the file.
The file is located in the root directory of the
CD-ROM.
7.
Follow the instructions given.
During the execution of the
.bat
file, you
may several times see a message asking you
whether existing files should be overwritten:
8.
Enter
A
to indicate that all files should be
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......