Installation
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
8-57
Procedure Details
2.
With a 6 mm drill, drill holes through the
wiremold and into the wall.
Note
. Take care to avoid any studs.
3.
Insert wall plugs into the holes drilled.
4.
Use the screws provided with the system
to secure wiremolds.
5.
Route the cable to the gantry inside the
mounted wiremold.
WARNING
Never
route the gantry-PC interface cable together with any other cables. Due
to safety and EMC/EMI considerations, the cable connecting the acquisition
PC to the gantry
must
be routed in separate wiremolds.
Procedure Details
6.
Use a part of the rubber protector supplied
with the system to protect the cable from
the wall to the gantry.
3 m (10 ft) of rubber cable protectors are
provided to cover cables running from the
gantry to the wall (power and network cables)
and the cable between gantry and PC. The
cable protectors should be applied where
traffic across cables must be expected.
7.
Cut cable protectors to desired lengths.
Open the membrane at the base of the
protector to slide in the cable.
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......