Installation
8-64
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
Procedure Details
13.
Attach the eye terminal of the
yellow/green ground wire to the ground
screw on the power supply frame.
14.
Remove the rear connector panel for the
ECG gate by unscrewing the four M4 x 8
screws holding the panel.
15.
Unscrew the two M3 screws to remove the
small bracket for the Ethernet connector
16.
Place the bracket on the Ethernet
connector and mount the connector to the
rear panel.
17.
Mount the panel again on the rear of the
table console.
Note
. Strain relief should be added to the gantry-PC interface cable as follows:
Procedure Details
18.
After making all connections, take all
excess cable length to the interior of the
gantry.
Gound wire
M3 screws
Bracket