Installation
8-26
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
Procedure Details
The cart-mount option is shipped in a plywood
box, 31.5
×
47
×
37 inch / 80
×
120
×
93 cm.
The wall-mount and desk-mount options are
shipped in cardboard boxes, 19.7
×
110.6
×
10.6 inch / 50
×
281
×
27 cm.
8.
Open the boxes and remove the protection
bag by carefully cutting around the edges.
Note
. Check the packing list and make sure that all parts are present.
8.3.3.2
Removing End bars and Shipping Brackets
Procedure Details
In order to be able to mount the jack bars, the
extensions at both ends of the pallet must be
removed.
1.
Use a Philips bit to remove all 4 steel
angles, and remove the 2 pieces of wood.
2.
Using a 16 mm wrench, dismount the 2
small ‘Z’ shipping brackets, and push the
bolts down.
3.
Cut the nylon strap and remove the 2
wedges.
Remove bolts
Cut strap
Remove wedges