CardioMD AC Option
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
12-77
12.5.3.10
Mounting Line Source Compartment Covers
This section explains how to install line source compartment covers:
•
While the transmission scanner is mounted on the detector, and
•
When the transmission scanner is dismounted and the line source compartment has been
detached from the AC option bracket.
Transmission Scanner Mounted on Detector
Procedure Details
1.
If required, use the hand controller to
position the detector assembly
conveniently for access to the transmission
scanner you are going to work with.
2.
If required, use the crank supplied with the
CardioMD system to position the line
source compartment conveniently.
3.
Insert the service cover at the end of the
line source compartment.
The small bolts mounted at the corners of the
cover must fit into the groove in the line
source compartment.