CardioMD AC Option
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
12-53
12.5.3.5
Replacing the AC Option Servo Motor
Procedure Details
1.
Use a T08 Torx key to remove the two
collision sensors above the motor Acme
screw assembly.
The arrows in the photo indicate the collision
sensors to be removed.
2.
Follow the procedure given in the section
Removing the Line Source Housing
, page
12-68 to remove the line source housing.
WARNING
Make sure that the transportation screws are in place and locking the shutter
before attempting to remove the source holder. Failing to lock the shutter may
cause you to be exposed to radiation from the source inside the holder.
Due to the Gd-153 source, the CardioMD line source housings contain
significantly higher levels of radiation than normally used in clinical imaging.
Potential for radiation exposure exists. Use good radiation safety practices.
Handling of line source housings with Gd-153 sources installed shall only be
performed by Philips Nuclear Medicine employees that have had radiation
safety training classes and trained on proper handling of radioactive material.
CAUTION
RADIOACTIVE
MATERIAL
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......