Installation
12-102
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
Procedure Details
1.
Place a lock washer over each Allen
screw, concave side of washer facing the
screw head.
Caution
. Make sure to place one lock washer
over each Allen screw. Leaving out the washer
entails a risk of damaging the detector.
2.
Mount the 3 guide pins on the end piece of
each detector, taking care to mount the
coded guide pin towards the top of the
detector (where the EDC board is located).
The photo on the right shows detector 1 with
covers mounted.
Note
. One of the guide pins for each detector
is coded and must be mounted at the top of the
detector, with its flat side facing
away
from
the table console. By doing this, you ensure
that the correct transmission scanner is placed
on each detector:
•
Transmission scanner 1 must be mounted
on detector 1 (facing detector 2), and
•
Transmission scanner 2 must be mounted
on detector 2 (facing detector 1).
This side away from
table console
Coded guide pin
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......