Calibration
12-28
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
Figure 12.23 Gantry position for detector check and calibration with the AC option mounted
Instructions for using the AC option point source holder are provided in Chapter 7
CardioMD AC
Option
of the CardioMD Operator’s Manual.
Gantry and source position for PMT update
and intrinsic QC check of detector 2
1
2
X = 10
Y = 1190
Detector angle = – 31°
Radiation
source
1
2
X = 10
Y = 1190
Detector angle = – 59°
Gantry and source position for PMT update
and intrinsic QC check of detector 1
Radiation
source