Installation
8-52
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
Procedure Details
10.
Move the arm and washers into alignment
with the wall mounting plate rings.
11.
Insert the pivot shaft so that the slot in the
pivot shaft allows the cable to escape the
arm/shaft/mounting assembly.
12.
Replace the M6
×
20 cap screw at the rear
of the arm.
13.
Remove the top cover from the arm.
14.
Replace the cable grommet at the bottom
of the pivot shaft.