Calibration
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
4-21
Procedure Details
The ECor page appears.
7.
Unless a major repair requires a calibra-
tion from scratch, select Update ECOR ta-
ble.
8.
Select the appropriate isotope (either Tc-
99m or Co-57).
9.
Select detector 1 or 2 to be calibrated.
Note
. If updating ECor, either isotope can be
selected. The update is not dependent upon the
isotope used during the initial ECor calibra-
tions.
10.
Click Continue.
11.
Press and hold the hand controller Colli-
sion Override button for a minimum of 4
seconds before proceeding as follows:
12.
While still pressing Collision Override,
click OK to accept automated motions
bringing the system into position for cali-
bration of the selected detector.
Note
. The hand controller Collision Override
button must be kept depressed during all
automated motions to enable power to the
motors.
Note
. Automated motions must be started
from the patient load position.
13.
Place the point source holder at a distance
of approximately 1.85 meters (6 ft.) from
the detector to be calibrated.
14.
Place a Tc-99m or Co-57 point source of
approximately 40 MBq or 1 mCi in the
holder.
15.
Make sure that the entire detector FOV is
exposed to the source and that the count
rate is < 40 kcps.
1
2
75
+/
- 3
c
m
185 +/- 3 cm
1
2
75
+/
- 2
c
m
185 +/- 3 cm
16.
When the source is in position, click Start
to begin the calibration.
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......