Installation
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
8-67
8.3.3.25
Removing Detector Insulation
The system must stabilize to room temperature before the detector insulation can be removed.
Otherwise, there is a risk of damage to the crystal.
WARNING
The thermal protection must never be removed before the detectors have been
in a controlled environment and the detectors have reached room temperature.
It is suggested to allow the detector temperature to stabilize for a period of 24
hours with the thermal protection on the detector if the equipment is being
relocated from one temperature to an extremely different temperature
environment.
The accepted rate of change should be less than 15° F (8° C) per hour. Do not
locally heat or cool the detector so as to produce gradients greater than 5° F
(3° C) across the detector face. Failure to comply may result in detector
crystal breakage.
8.3.3.26
Mounting the Patient Table and Adjusting the Table Support
Procedure Details
1.
Use 8 pcs. M8 low-head screws to mount
the patient table pallet.
2.
Adjust the two feet under the table support
to allow the table pallet to rest on the
rollers.
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......