Installation
8-66
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
8.3.3.24
Tightening the Rear Anchor Bolts and Testing Motion
Procedure Details
1.
Use the hand controller to move the
detectors up and in towards the patient
table, leaving the x-axis approximately 8”
(20 cm) from the rear of the table.
2.
Tighten the 2 rear anchor screws using a
10 mm Allen key.
3.
Using the hand controller, check that the
system will move along all axes.
4.
Leave the detector assembly in the top
position in order to be able to remove the
detector insulation.