CardioMD AC Option
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
12-65
Procedure Details
32.
Assemble the spring and end stop as
shown in the photo.
33.
Attach the spring and end stop to the
collision detection shutter end cover.
34.
Mount the collision detection shutter end
cover onto the line source compartment.
Remember to mount the nut on the inside of
the line source compartment behind the spring
and end stop.
Proceed to remount covers, assemble the
transmission scanner and install it on the
detector:
35.
Install the cable connecting the line source
compartment to the bracket by reversing
instructions provided in section 12.5.3.2,
page 12-43.
36.
Install line source compartment covers
(see page 12-79).
37.
Attach the line source compartment to the
bracket by reversing instructions in section
12.5.1.2, page 12-33.
38.
Mount the transmission scanner onto the
detector. See section 12.6.3.4, page 12-
103.
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......