Repair Procedures
12-64
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
Procedure Details
27.
Mount the spring onto the actuator
bracket.
28.
Check the alignment of optocoupler
detecting the position of the shutter
indicator / manual lever.
When the shutter is in the closed position
(shutter indicator / manual lever in the IN
position), the position must be as indicated on
the photo.
29.
If required, the position of shutter
indicator / manual lever can be adjusted by
loosening the two Allen screws shown.
30.
Finally, operate the manual lever a few
times and check that it is still moving
smoothly.
31.
Reconnect the solenoid cable.