Installation
8-68
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
8.3.3.27
Checking Motion Limits
Follow instructions provided in Chapter 4
Calibration
to check that the motion limits are properly
set. If not, calibrate the motion limits, again following instructions given in Chapter 4.
8.3.3.28
Mounting Stays for Base Cover
To mount the four stays that will support the gantry base cover, proceed as follows:
Procedure Details
1.
Locate the four stays for the gantry base
cover.
2.
Mount the stays to the gantry frame.
Note
. Do not use excessive force when
tightening the stays as they may break if
subjected to too large a torque.
8.3.3.29
Mounting Covers
Procedure Details
The cover of the tower must be removed
before the base cover can be mounted.
1.
Dismount the two screws at the rear: one
at the bottom and one at the top of the
cover.
2.
Remove the tower top cover (the ‘hat’).
3.
Remove the two screws on the tower’s
front side, again, one at the top and one at
the bottom.
4.
Remove the cover.
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......