Detailed Description
12-12
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
12.3.3
AC Option Bracket
The AC option bracket attaches the AC option to the end of the detector casting and it contains the
drive that moves the line source compartment in and out.
Figure 12.9 The AC option bracket seen from the outside
The bracket housing is made of aluminum plate welded to form a rigid base for the support of the
complete AC option assembly.
12.3.3.1
AC Option Bracket Mechanics
Locking Mechanism
The locking mechanism consists of a steel plate and a handle. The steel plate is screwed onto the
AC option bracket and cutouts in the plate allow the bracket to be mounted onto the guide pins of
the detector housing. The handle locks the AC option in place.
Handle in open position
Handle in locked position
Figure 12.10 Bracket locking mechanism
AC option bracket
4 slide bearings
Rails
Line source compartment