1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
12-67
Procedure Details
4.
Unscrew the two wires from the collision
sensors to be replaced.
The connector pinout is shown to the right.
1
2
3
4
5
6
7
8
300
mm
45
0 m
m
740 mm
74
0
mm
5.
Mount the new set of collision sensors by
reversing the steps 1 to 4.
Note
. Sensors
must
be mounted square and
flush so that they mate with the cover surface
evenly across their surface.