Diagnostics
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
5-11
Channel status
Port Signal
Explanation
CHA_FAULT Rotate amp. fault
High (LED Off) during normal use
Low (LED On) when motor power is not present
CHA_LR Not
used
CHA_LL Not
used
CHA_BRK Not
used
CHA_ENA
Rotate amp. enable
Low (LED On) during normal use
High (LED On) when motor power is not present
CHB_ENA Not
used
CHB_BRK Not
used
CHB_LL Not
used
CHB_LR Not
used
CHB_FAULT Not
used
Table 5.4 Rotate motion controller channel status
5.3.4.2
Linear Motion Controller Port Assignments
The only signals connected to the input port of the linear motion controller are the hand controller
inputs. All buttons are low, (i.e., the LED is lit when the button is pressed). The Stop button is not
connected to the motion controller.
Input port
Port
Signal
IN00
HS Motion Enable
IN01 HS
Mark
IN02
HS X Out
IN03
HS X In
IN04 HS
Table
Out
IN05 HS
Table
In
IN06
HS Y Up
IN07
HS Y Down
IN08
HS Rotate CCW
IN09
HS Rotate CW
IN10
HS Start ACQ
IN11 HS
Collision
Override
IN12
IN13
IN14
IN15
IN16
IN17
IN18
IN19
Table 5.5 Linear motion controller input port
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......