Installation
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
8-33
8.3.3.7
Anchoring the Mounting Plates
Procedure Details
1.
Place the 2 mounting plates over the holes.
Make sure that the side with countersunk
around the M10 hole is facing upwards.
2.
Mount 4 of the 6 M12
×
30 anchor screws
and washers shipped with the system.
3.
Do
not
tighten the anchors yet.
Note
. Seismic installations require use of the optional CardioMD Seismic Kit. Instructions for
seismic installation are included with the kit.
8.3.3.8
Positioning the Gantry
Procedure Details
1.
Lower the jacks until the gantry is about
3/4” (20 mm) above the mounting plates.
If needed, push the gantry until the 4
screw heads below the gantry end bars
align with the 2 holes in each mounting
plate.
The screw heads will guide the gantry while
the jacks are being lowered the last 12 mm to
the floor.
2.
When the jacks are free, check that the
four M10 mounting screws can be
installed.
3.
Dismount the jack beams.
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......