Repair Procedures
12-50
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
Procedure Details
21.
Install the line source position indicator
arm, taking care to first fasten it to the
slide on the right, as indicated by the
arrow in the photo.
22.
Use the 2 mm Allen key and the 7 mm
fork wrench fasten the indicator arm and
the jam nut.
23.
Attach the green connector to the motion
controller.
24.
Mount the spring onto the end of the
indicator arm.
25.
Connect all remaining cables to the source
motion board.
26.
Assemble the collision sensor, paying
attention to insert the bracket correctly.
Note the orientation of the oblique edge in
the photo on the right.