Installation
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
8-43
Procedure Details
3.
Connect the hand-controller/LED/E-stop
cable to the DB25 connector on the
internal connector panel.
4.
Connect the power cable to the IEC
connector.
5.
Connect the FireWire cable to any of the
ports on the FireWire hub.
6.
Attach the eye terminal of the
yellow/green ground wire to the ground
screw on the power supply frame.
7.
Remove the rear connector panel for the
ECG gate by unscrewing the four M4 x 8
screws holding the panel.
8.
Unscrew the two M3 screws to remove the
small bracket for the Ethernet connector
DB25 connector
IEC connector
FireWire hub
Gound wire
M3 screws
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......