Installation
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
8-71
Procedure Details
16.
Attach the crank handle in the holders on
the rear side of the gantry base.
8.3.3.30
Setting Up the ECG Gate
Note
. The ECG gate is not shipped with the CardioMD system but is supplied separately by
Philips Medical Systems.
Procedure Details
1.
Unpack the ECG gate.
2.
Position the ECG gate and the dedicated
gate support bracket supplied with the
CardioMD system next to the table
console as shown in the photo on the right.
Note
. The support bracket will prevent the gate from being accidentally pushed into the detector
rotation path.