Installation
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
8-3
8.1
Installation Requirements
8.1.1
Introduction
This installation section describes the installation of CardioMD Series III system. Should you
need to install a CardioMD Series I or II system, refer to Chapter 8
Installation
of the service
manual originally provided with the CardioMD Series I or II system.
Note
. A few of the photos in this chapter actually show CardioMD Series I/II systems. This is the
case only where these details of the system are insignificant for the installation procedure.
Note
. Regardless of the system version, read the document
CardioMD Camera Site Planning
, part
no. 9347-0112.
Note
. Section 8.3.4 Software Setup on page 8-75 does provide some information on setting up
software under Windows 2000, which is used on Series I and II systems acquisition PC’s.
8.1.2
Ideal Floor Space Allocation
Some general considerations should be followed for all nuclear camera systems. These include
comfortable workspace, good visual access to patients on the equipment, sufficient service access,
and patient safety. All nuclear equipment contains components such as the sodium iodide crystals
and high speed electronics that require reasonably well controlled temperature, humidity, and
power.
Complete camera systems can be installed in a single room or with the gantry and acquisition
station in one room and the processing station in a separate room.
8.1.3
Minimum Floor Space Allocation
Minimum floor space layouts are shown for the CardioMD system including processing station
electronics.
The electronics do not require a separate environmentally controlled room, provided that the
proper cooling and humidity ranges are available.
# Equipment
Width
Depth
Height Weight
1a
Gantry and table assembly*
Incl. +/– 5 cm travel
112 inch
284 cm
58 inch
150 cm
58 inch
150 cm
1797 lbs.
815 kg
1b
Gantry and table assembly*
Incl. +/– 5 cm travel
83 inch
210 cm
58 inch
150 cm
58 inch
150 cm
1797 lbs.
815 kg
2 Collimator
cabinet
17 inch
43 cm
17 inch
43 cm
40 inch
102 cm
75 lbs.
33 kg
3a Cart-mounted
acquisition PC
25.2 inch
63.9 cm
25.4 inch
64.5 cm
47.3 inch
120 cm
66 lbs.
30 kg
3b
Wall-mounted acquisition PC
30.8 inch
79 cm
15.2 inch
39 cm
22.6 inch
58 cm
51 lbs.
23 kg
3c Desk-mounted
acquisition PC
22.8 inch
58 cm
15.0 inch
38 cm
11.8 inch
30 cm
10 lbs.
5 kg
* The CardioMD system is available in a standard configuration (1a) and a reduced room size version (1b)
Table 8.1 Minimum floor space allocation for the CardioMD system
I, II
III
I, II
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......