Spare Parts
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
9-37
Procedure Details
4.
Attach the collimator number label (I or
II) onto the collimator in the location
indicated in the diagram on the right.
5.
After replacing a collimator, you must
acquire:
-
A new center of rotation correction
table
-
New extrinsic uniformity calibration
tables.
Refer to the Operator’s Manual Chapter 5
Maintenance
for procedures for performing
center of rotation calibration and extrinsic
uniformity calibration.
I or II label position
Slanted part
Handle
Side towards
patient