Repair Procedures
12-52
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
Procedure Details
instructions provided in section 12.5.3.2,
page 12-43.
32.
Install line source compartment covers
(see page 12-79).
33.
Attach the line source compartment to the
bracket by reversing instructions in section
12.5.1.2, page 12-33.
34.
Mount the transmission scanner onto the
detector. See section 12.6.3.4, page 12-
103.
12.5.3.4
Replacing the Motion Controller
The hardware of the motion controllers of the two transmission scanners is identical but the
motion controller firmware is different, according to whether the transmission scanner is for
detector 1 or for detector 2. Thus, the motion controllers cannot be swapped.
Note
. Motion controller 9EMC1411 must be installed in transmission scanner 1 and motion
controller 9EMC1410 must be in transmission scanner 2.
The procedure for replacing any of the two motion controllers is described below.
Procedure Details
1.
Follow the procedure for replacing the AC
source motion board outlined in section
12.5.3.3 starting page 12-44 to dismount
the motion controller and source motion
board assembly.
2.
Use a T6 Torx key to unscrew the motion
controller from its mounting plate.
The arrows in the photo on the right indicate
the position of the screws.
3.
Follow instructions in section 12.5.3.3 to
reassemble and mount the transmission
scanner.
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......