Repair Procedures
12-34
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
12.5.2
AC Option Bracket Repairs
12.5.2.1
Dismounting Detector Cable
The 9CBL1357 cable connecting the bracket to the detector is easily replaced when the line source
compartment has been dismounted from the bracket.
12.5.2.2
Disassembling the Bracket
To further disassemble the bracket for repairs, proceed as follows:
Procedure Details
1.
Use a Torx T20 key to unscrew the two
torx screws in bottom side of the bracket.
2.
On the rear of the bracket, use a 4 mm
Allen key to unscrew the 7 screws holding
the steel mounting plate to the bracket.
Note
. Note the straight edge in the cutout next
to the handle which is used to code the
transmission scanner for detector 1 or detector
2. The photo shows coding for detector 2. The
plate can be flipped to code the scanner for
detector 1.