Software Update Procedures
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
7-7
Procedure Details
responds to commands in the ttermpro emula-
tor. This is normal.
Now the ECG connection must be restored
and the EDC module must be power cycled:
7.
Remove the serial interface cable from the
connector A1 RS232 and reconnect the
ECG cable (see step 2 in the section
Preparing for Firmware Upgrade
on page
7-4).
8.
Remove the EDC power cable connected
to connector A4 POWER. Reconnect the
power cable.
Now proceed to upgrade the logic firmware of
the other EDC board:
9.
Connect the serial interface cable to the
other EDC board as described on page 7-4.
10.
Repeat steps 1 to 8 of this procedure to
upgrade the logic firmware of the EDC
module on the other detector.
11.
When both EDC modules have been up-
dated, remount the detector top cover fol-
lowing instructions provided in Chapter 6.
Repair Procedures
.
7.2.3.3
Upgrading EDC Modules 9CSY0900-B02 or higher
Before starting the installation, check the release note on the CD-ROM. You find the release note
in the directory
Release notes
. The release note may provide information that is more recent
than this manual.
Note
. If you are using a laptop acquisition PC, make sure that you will be using either COM1 or
COM2 to communicate, since the software used for downloading only supports these ports. See
section 7.2.2 on page 7-2 for instructions on determining the COM port number.
Connector A4 POWER
on the EDC board
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......