CardioMD AC Option
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
12-99
12.6.3
Mechanical Installation
12.6.3.1
Preparing Detector Covers
Procedure Details
1.
Remove the pre-cut knock-out pieces from
both ends of the detector top cover, using
an appropriate pair of pliers.
Caution
. Be careful when removing the
knock-out pieces. Using excessive force may
damage other parts of the cover.
2.
Dismount the two detector end covers.
3.
If necessary, use a small hammer to
remove the three pre-cut knock-out pieces
in both covers.
4.
Take care not to damage the detector
cover outside of the knock-out pieces.