Acquisition PC
3-32
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
3.4.2
Laptop Acquisition PC
Series III CardioMD systems are supplied with a laptop acquisition PC with the following
minimum specifications:
IBM compatible, 1.5 GHz (Pentium M), 8 Mbyte CPU RAM, 256 Mb SDRAM PC100, 40 Gbyte
hard disk, DVD-CD R/W drive, 32 Mb Graphics (1024
×
768), EtherNet adapter (PCI
10/100BTX), IEEE1394 FireWire interface, 2
×
USB 2.0, 15” TFT monitor.
The PC software operates under MicroSoft Windows XP.
3.4.3
PC Connections
The acquisition PC must be powered with 115 V from the gantry.
On Series I and II systems, acquisition PC signal connections are the following:
•
FireWire. The cable between the acquisition PC and the gantry contains a FireWire cable,
which is connected to the PC’s FireWire interface and to the Firewire connection on the
gantry connector panel.
•
Local area network (ethernet). A standard network cable with RJ45 connectors connects the
PC to the local area network. The cable between the acquisition PC and the gantry contains
the network cable, which ends in an RJ45 female connector near the gantry connector panel.
•
E-Stop. A two-wire cable connecting the Emergency Stop button located next to the
acquisition PC to the gantry.
On Series III systems, acquisition PC signal connections are the same (power, FireWire, LAN, E-
Stop), plus signals for the Collision, Power OK and E-Stop LED’s next to the laptop PC.
However, all connections are internal to the gantry.
The gantry must be connected to the local area network via the connector on the rear panel of the
table console (see Figure 3.12, page 3-17).
I, II
III
III
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......