CardioMD AC Option
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
12-31
Procedure Details
The contents of the
ACoption.ini
file
determine whether the AC option is active or
not. The arrow in the figure on the right
indicates the entry to look for.
15.
To disable the CardioMD AC Option, edit
the entry to say
ACoption=0
.
This setting allows the CardioMD system to
function without the AC option.
16.
Select
File
,
Save
to save the modified file.
17.
Click the Close button to close down the
Notepad program.
Powering Down the Gantry
Procedure Details
Close down the acquisition PC and power off
the CardioMD gantry:
18.
Follow the system shutdown procedure
provided in Chapter 3
Getting Started
of
the CardioMD Operator’s Manual.
Detaching the Transmission Scanner
Procedure Details
19.
Unscrew and disconnect the cable to the
transmission scanner from the socket on
the detector.
CardioMD configuration file – AC option disabled