Installation
8-24
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
8.3.3
Installation Procedure
8.3.3.1
Unpacking
Move the crate as close to the place of installation as the access permits, still having workspace to
unload around it.
Figure 8.16 Minimum workspace and corridor width
Procedure Details
The gantry is shipped in a wooden crate (34
×
61
×
63 inch / 86
×
155
×
160 cm).
1.
Remove the plywood sides and top of the
crate.
2.
Remove the protection bag by carefully
cutting around the edges.
Note
. Leave the detector insulation in place at this point.
Required work space for unpacking and
unloading: 100 x 92 inch (256 x 236 cm)
If corridors have a 90
°
corner, the corridor must be at
least 67 inch (175 cm) wide