CardioMD AC Option
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
12-101
12.6.3.2
Mounting Guide Pins on Detectors
Three guide pins must be mounted on the end piece of each detector. These guide pins fit into
sockets in the bracket of the transmission scanner and carry the transmission scanner when it is
mounted on the detector.
Procedure Details
The photo on the right shows the detector end
piece without the cover.
The CardioMD AC Option comprises 2 sets of
3 guide pins, complete with Allen screws and
lock washers, to be mounted on the end piece
of each detector.
The photo on the right shows the guide pins
for one detector.