Diagnostics
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
5-7
COLLIS. OK
This green LED will be on during normal operation. During collision the LED is
turned off.
COLLISION
This red LED is on during collision and off during normal operation.
COLLIS.
DISABLED
This red LED is on when the Collision Override switch is on (position down) to
warn that the collision circuitry is disabled.
ULT_STOP
DISABLED
This red LED is on when the Ultimate Stop Override switch is on (position down)
to warn that the ultimate stops are disabled.
ROT.
ULT_OK
This green LED is on during normal operation. The LED is turned off when the
ultimate stop switch on the rotational motion has been triggered (motion limit
exceeded), or when either the HORIZ. ULT_OK or VERT. ULT_OK has been
triggered.
HORIZ.
ULT_OK
This green LED is on during normal operation. The LED is turned off when the
ultimate stop switch on the horizontal (X) motion has been triggered (motion limit
exceeded), or when VERT. ULT_OK has been triggered.
VERT.
ULT_OK
This green LED is on during normal operation. The LED is turned off when the
ultimate stop switch on the vertical (Y) motion has been triggered (motion limit
exceeded).
E_STOP OK
This green LED is on during normal operation. The LED is turned off when one of
the Emergency Stop buttons has been pressed.
TABLE
E_STOP ON
This red LED is off during normal operation. The LED is turned on when the
gantry E-Stop button has been activated.
REMOTE
E_STOP ON
This red LED is off during normal operation. The LED is turned on when the E-
Stop button next to the acquisition PC has been activated.
RUNAWAY
This red LED is off during normal operation. The LED is turned on when one of
the motion controllers has issued a Runaway signal.
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......