Setting Up FTP and Telnet
5-52
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
Procedure Details
10.
Close the Add/Remove Programs dialog
by clicking the Close button in the upper
right-hand corner of the window.
11.
In the Control Panel, double-click
Administrative Tools.
12.
In the Administrative Tools window,
double-click Internet Services Manager.
13.
In the Internet Information Services
window, click the + next to the name of
the computer to unfold the branch of the
tree.
14.
Right-click Default FTP Site and select
Properties from the drop-down menu.
15.
In the Default FTP Site Properties dialog,
select the Home Directory tab at the top of
the window.
16.
On the Home Directory page, set Local
Path to
C:\
.
17.
Click OK.