Repair Procedures
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
6-7
6.3.4
Replacing Detector Power Supply and EDC Board
Procedure Details
These units are located directly under the
detector top cover.
1.
Remove the top cover and disconnect all
cables from the unit to be removed.
To remove the EDC module:
2.
Unscrew the 4 screws attaching the EDC
module to its support plate.
3.
Set the detector selector switch to position
DET1 or DET2, depending on the detector
at which the module is mounted.
After mounting the new EDC module, check
whether it has the correct software:
4.
Power up the CardioMD system and the
acquisition PC.
5.
Open the latest log file in the acquisition
PC directory
C:\cardiocam\SystemLog
.
6.
Check that both EDC modules have the
same software version.
See Chapter 7
Software Update Procedures
,
the section
Checking Installed Software
Versions
, for detailed instructions.
7.
If they do not, update EDC software.
See Chapter 7
Software Update Procedures
for instructions.
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......